• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL CARDIOMEMS HOSPITAL ELECTRONICS SYSTEM; SYSTEM, HEMODYNAMIC, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL CARDIOMEMS HOSPITAL ELECTRONICS SYSTEM; SYSTEM, HEMODYNAMIC, IMPLANTABLE Back to Search Results
Model Number CM3000
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/12/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device is included in the error 5 advisory notice issued by abbott on 01 june 2018.The investigation codes along with investigation results will be provided in a subsequent submission.
 
Event or Problem Description
It was reported that the hospital electronics unit received an error 5 message, which indicates an issue related to temperature and barometric pressure.Troubleshooting the unit was able to resolve the issue.
 
Additional Manufacturer Narrative
The reported issue of "error 5" was confirmed.Backend verification confirmed that the error 5 issue was present on 12sep2022.The affected unit was operating on software version 1.7.1-r5 at the time of complaint.The reoccurrence of error 5 notification was due to a software issue with software version 1.7.1-r5.The error would not clear on its own, and the error was required to be manually cleared.The issue was corrected under the latest software version.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOMEMS HOSPITAL ELECTRONICS SYSTEM
Common Device Name
SYSTEM, HEMODYNAMIC, IMPLANTABLE
Manufacturer (Section D)
ST. JUDE MEDICAL
387 technology circle nw suite 500
atlanta GA 30313
Manufacturer (Section G)
ST. JUDE MEDICAL
387 technology circle nw suite 500
atlanta GA 30313
Manufacturer Contact
bob fryc
one st. jude medical drive
st. paul, MN 55117
6517565400
MDR Report Key15542569
Report Number3004936110-2022-00515
Device Sequence Number6575686
Product Code MOM
UDI-Device Identifier05415067029331
UDI-Public05415067029331
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P100045
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 10/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCM3000
Device Catalogue NumberCM3000
Device Lot Number6268623
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/12/2022
Supplement Date Received by Manufacturer10/24/2022
Initial Report FDA Received Date10/05/2022
Supplement Report FDA Received Date10/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/09/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
3004936110-06/01/18-001R
Patient Sequence Number1
Patient SexUnknown
-
-