| Model Number |
406701 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
07/06/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Upon removal of the introducer, a portion of the device detached and remained in the patient when suturing the insertion site.The detached portion of sheath was noted on fluoroscopy.The site was re-accessed to allow for snare insertion and retrieval of the sheath portion which had migrated toward the right ventricular region.The snare was used to retrieve the detached portion with no apparent injury to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported device detachment remains unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected data: h2, h6.
|
| |
|