| Model Number |
04.503.740 |
| Medical Device Problem Codes |
Break (1069); Material Integrity Problem (2978)
|
| Health Effect - Clinical Codes |
Unspecified Tissue Injury (4559); Insufficient Information (4580)
|
| Date of Event |
09/09/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported (b)(6) 2022 the surgeon removed the implants from the first surgery as the implants were damaged inside the patient.No information is known on patient status.This is report 1 of 3 for (b)(4).This report is for ti matrixmandible 7x23 angle recon pl right 2.5mm thick.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.Part: 04.503.740, lot: 61p1351, release to warehouse date: 02.July.2020, expiration date : na, supplier: (b)(4), manufacturing site: werk selzach.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that matrixmand reco-pl angl r 7+23ho t2.5 ti was broken along the shaft into several pieces.Only two pieces were returned.Scratches and discoloration, which could have been a result of explantation, were also observed on the surface of the returned broken plate.No other issues were noted with the device.A dimensional inspection for thematrixmand reco-pl angl r 7+23ho t2.5 ti was unable to be performed due to post manufacturing damage.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the matrixmand reco-pl angl r 7+23ho t2.5 ti would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.If information is obtained that was not available for this medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|