This is filed to report damage to the soft tip of the device and presence of thrombus.It was reported that on (b)(6) 2022, a patient presented with grade 4 degenerative mitral regurgitation (mr), history of aortic valve replacement, and a watchman left atrial appendage implant.During the procedure with a steerable guide catheter (sgc), the device was unable to cross the septum.Force was applied in effort to cross, but the sgc could not advance.The sgc was removed and damage was noted on the soft tip, as the surface was not smooth and appeared to be dented.The soft tip was fully intact.There was also a clot observed on the tip of the sgc.Additional heparin was administered and no additional treatments were needed.There was no adverse patient sequela.A guide catheter balloon was used on the septum, and a new sgc was able to pass successfully.The first xtw clip delivery system (cds) was advanced, and the shaft appeared to have an unusual bend while checking the trajectory of the clip.The cds curved to the medial side.There was speculation that the curve was due to the anatomy (aortic valve replacement).As a precautionary measure, the cds was removed and replaced.The clip was deployed and mr was reduced to grade 2.There were no adverse patient effects or clinically significant delays caused by any of the abbott devices.No additional information was provided.
|
The device was not returned for analysis.A review of the lot history identified no manufacturing nonconformities issued to the reported lot that would have contributed to the reported event.Additionally, a review of the complaint history did not indicate a lot-specific product issue.The investigation determined the reported failure to advance appears to be related to patient anatomy.The dented soft tip was due to the failure to advance.Thrombosis is listed in the instructions for use as known a possible complication associated with mitraclip procedures.The reported unexpected medical intervention was a result of case-specific circumstance.There is no indication of a product issue with respect to manufacture, design, or labeling.
|