| Model Number |
D010 |
| Medical Device Problem Codes |
Premature Discharge of Battery (1057); Delayed Charge Time (2586); Defective Device (2588); Battery Problem (2885)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/27/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) exhibited device longevity from one and a half years to elective replacement indicator (eri) in a span of six months.Data analysis determined that the device is depleting normally.Eri was declared due to two high voltage charges in excess of fifteen seconds.The device was explanted.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental report will be filed.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has been received for analysis and upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a medwatch report will be filed.Supplemental correction to correct d6b explant date.
|
| |
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) exhibited device longevity from one and a half years to elective replacement indicator (eri) in a span of six months.Data analysis determined that the device is depleting normally.Eri was declared due to two high voltage charges in excess of fifteen seconds.The device was explanted.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has been received for analysis and upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a medwatch report will be filed.Correction to h6: included device code for delayed charge times.
|
| |
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) exhibited device longevity from one and a half years to elective replacement indicator (eri) in a span of six months.Data analysis determined that the device is depleting normally.Eri was declared due to two high voltage charges in excess of fifteen seconds.The device was explanted.No additional adverse patient effects were reported.This icd was returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has been received for analysis and upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a medwatch report will be filed.Supplemental correction to correct d6b explant date.Correction to h6: included device code for delayed charge times.The returned implantable cardioverter defibrillator (icd) was analyzed and an engineering-level longevity prediction calculation was completed to assess the rate of battery depletion.Given the programmed parameters and other data stored within the memory of the device, the results of this calculation indicated that the actual rate of battery depletion fell within an acceptable range.Next, a series of diagnostic tests were conducted that verified the performance of pacing, sensing, defibrillation, and recording functions.Having met the engineering longevity prediction, functionally passed all returned product testing, and with no further information to indicate a product performance issue, we have concluded that this device experienced normal battery depletion.
|
| |
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) exhibited device longevity from one and a half years to elective replacement indicator (eri) in a span of six months.Data analysis determined that the device is depleting normally.Eri was declared due to two high voltage charges in excess of fifteen seconds.The device was explanted.No additional adverse patient effects were reported.This icd was returned for analysis.
|
| |
|