|
Model Number 97715 |
Device Problems
Overheating of Device (1437); Connection Problem (2900); Insufficient Information (3190); Appropriate Term/Code Not Available (3191)
|
Patient Problems
Burn(s) (1757); Pain (1994); Burning Sensation (2146)
|
Event Date 09/19/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Concomitant medical products: product id: 97755, serial# (b)(4), product type: recharger.Other relevant device(s) are: product id: 97755, serial/lot #: (b)(4), udi#: (b)(4).Date is approximate.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from a patient (pt) regarding an external device.It was reported that for the past couple of weeks, they started to notice that the rubber coating was getting pushed up where the recharger (rtm) cord entered the paddle, and they could see the wire within.Pt then said over the weekend they were charging, and it felt like something had burnt them, so they kept moving the rtm and finally they took the rtm off to look at it because the feeling was irritating them.Pt said when they took the rtm off and looked at the cord, it was "arching fire".Pt said they hurt for a few days after that.A replacement rtm was requested.
|
|
Event Description
|
Additional information was received from the patient (pt).It was reported that the circumstances that led to the recharger (rtm) wire being exposed and burning the pt were that the black part on the rtm wire came up and you could see the wire.The patient felt pain and burning so they moved it over on their hip.It got worse, so they took it off their hip and they could see fire coming from it.They don't know why it was exposed.The steps taken to resolve the issue were that they were sent a new one.The replacement resolved the issues.
|
|
Manufacturer Narrative
|
Continuation of d10: product id: 97755, serial# (b)(6), product type: recharger.H3: analysis of the recharger (serial# (b)(6)) found that the device was scrapped due to cable insulation being pulled out from the donut and due to wires being exposed and broken.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Continuation of d10: product id 97755 lot# serial# (b)(6) product type recharger medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|