• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER TRAUMA KIEL NAIL HOLDING SCREW FEMUR RETROGRADE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER TRAUMA KIEL NAIL HOLDING SCREW FEMUR RETROGRADE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 2353-3904
Medical Device Problem Code Device Difficult to Setup or Prepare (1487)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 09/10/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.
 
Event or Problem Description
It was reported that during the implantation of a right t2 alpha retrograde nail, the nail came off the jig, resulting in stripped threads.A second instrument was obtained from another hospital.Surgery was completed successfully with a delay of approximately 150 minutes.
 
Additional Manufacturer Narrative
Please note the correction to d9/h3 as the device was not returned for evaluation.The reported event could not be confirmed, since the device was not returned for evaluation and no other evidence was provided.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.A review of the device history was not possible because the lot number was not communicated.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.In case substantive information will become available in future that suggests otherwise we reserve the right to reopen the case.H3 other text : device disposition unknown.
 
Event or Problem Description
It was reported that during the implantation of a right t2 alpha retrograde nail, the nail came off the jig, resulting in stripped threads.A second instrument was obtained from another hospital.Surgery was completed successfully with a delay of approximately 150 minutes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAIL HOLDING SCREW FEMUR RETROGRADE
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM  D-24232
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key15543516
Report Number0009610622-2022-00452
Device Sequence Number5593329
Product Code HSB
UDI-Device Identifier07613327381825
UDI-Public07613327381825
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K203819
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2353-3904
Device Catalogue Number23533904
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/10/2022
Supplement Date Received by Manufacturer11/03/2022
Initial Report FDA Received Date10/05/2022
Supplement Report FDA Received Date11/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-