• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. JRNY II UNI MED XLPE INSERT SZ 7-8 12MM; PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. JRNY II UNI MED XLPE INSERT SZ 7-8 12MM; PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 74026182
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Post Operative Wound Infection (2446)
Date of Event 09/13/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that, after tka performed on unknown date, the patient experienced hematogenous infection of streptococcus after jrny ii uni med xlpe insert sz 7-8 12mm implanted, required a revision surgery.Washing and insert exchange was performed.The current health status of patient is unknown.
 
Additional Manufacturer Narrative
The device was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, per complaint details, an insert exchange (revision) with ¿washing¿ was performed due to a hematogenous infection of streptococcus after an unknown length of time post uni medial insert implantation.It was communicated that the requested clinical documentation was not available; therefore, additional contributing clinical factors could not be concluded, and it also cannot be concluded that there was a malperformance of the s+n component(s).With the limited information provided, the patient impact beyond the reported infection and subsequent revision with ¿washing¿ cannot be determined.No further medical assessment can be rendered at this time.A review of complaint history for the previous 12 months did not reveal similar events for the listed device.A review of the instructions for use documents for unicompartimental knee systems revealed that acute post-surgical wound infection has been identified as an adverse reaction, also active, local infection or previous intra-articular infections are identified as contraindications for unicompartimental knee replacement.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Factors that could contribute to the reported event include loss of sterility during procedure, patient reaction, and post-operative healing issue.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JRNY II UNI MED XLPE INSERT SZ 7-8 12MM
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15544136
Report Number1020279-2022-04318
Device Sequence Number5861972
Product Code HSX
UDI-Device Identifier00885556677247
UDI-Public00885556677247
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K190085
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Other,Foreign,Consumer,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number74026182
Device Catalogue Number74026182
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/14/2022
Supplement Date Received by Manufacturer11/03/2022
Initial Report FDA Received Date10/05/2022
Supplement Report FDA Received Date11/07/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-