| Catalog Number |
07475918190 |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/12/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The samples had been initially tested immediately after thawing.The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
There was an allegation of questionable low igf-1 elecsys results from cobas e 801 analytical unit serial number (b)(4) that did not match the clinical picture of the patients.One example was provided of an initial result of 80.2 ng/ml and a repeat result of 149 ng/ml.The questionable results were reported outside of the laboratory and were questioned by the physician.The repeat results fit the patients' histories and were considered valid.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation did not identify a product problem.The cause of the event could not be determined.The event was consistent with a preanalytical issue as the problem appeared to be solved when the samples were not frozen but just cooled.The calibration and qc data do not point to a reagent or instrument problem.
|
| |
|