• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. LGN CONE IMPACTOR HANDLE; PROSTHESIS,KNEE,PATELLOFEMOROTIBIAL,SEMICONSTRAINED,CEMENTED,POLYMERMETALPOLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. LGN CONE IMPACTOR HANDLE; PROSTHESIS,KNEE,PATELLOFEMOROTIBIAL,SEMICONSTRAINED,CEMENTED,POLYMERMETALPOLYMER Back to Search Results
Model Number 71442011
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/15/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that, during inspection, the tip of the cone broach handle of a lgn cone impactor handle was broken off.It wasn¿t able to correctly attach cone broaches.Surgery was performed, without any delay, with a s+n back-up device instead.Since this happened before procedure, patient was not yet involved.
 
Additional Manufacturer Narrative
H3, h6: the associated device was returned and evaluated.A visual inspection of the returned device confirms a piece that holds the rasp and trial broke off.The broken piece was not returned with the device.The device shows signs of significant wear and use.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.As the device broke and it can no longer fit its purpose, the contribution of the device to the reported event could be corroborated.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LGN CONE IMPACTOR HANDLE
Common Device Name
PROSTHESIS,KNEE,PATELLOFEMOROTIBIAL,SEMICONSTRAINED,CEMENTED,POLYMERMETALPOLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15544242
Report Number1020279-2022-04323
Device Sequence Number17786050
Product Code JWH
UDI-Device Identifier00885556537152
UDI-Public00885556537152
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K162775
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71442011
Device Catalogue Number71442011
Device Lot Number18BSC0012
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/14/2022
Supplement Date Received by Manufacturer10/07/2022
Initial Report FDA Received Date10/05/2022
Supplement Report FDA Received Date10/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/02/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-