• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON, S.A. BD PHASEAL¿ PROTECTOR (P55); INTRAVASCULAR ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON, S.A. BD PHASEAL¿ PROTECTOR (P55); INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Catalog Number 515117
Medical Device Problem Codes Device Contamination with Chemical or Other Material (2944); Material Protrusion/Extrusion (2979)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/09/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that the bd phaseal¿ protector (p55) had a small, sharp, plastic protrusion on the spike that fell into the bottle during the puncture.The sharp molding defect event occurred with 30 protectors.The following information was provided by the initial reporter, translated from german: "the piercing spike has a small plastic tip that enters the bottle during piercing.Customer informs us that when using ari.515117, it was noticed that a small plastic tip protrudes from the puncture spike, which gets into the bottle when puncturing.As a result, about 100 protectors were opened, of which a good 30 are affected.The batch has now been put to one side.It seems that the puncture spike was not designed properly during production.".
 
Event or Problem Description
It was reported that the bd phaseal¿ protector (p55) had a small, sharp, plastic protrusion on the spike that fell into the bottle during the puncture.The sharp molding defect event occurred with 30 protectors.The following information was provided by the initial reporter, translated from german: "the piercing spike has a small plastic tip that enters the bottle during piercing.Customer informs us that when using ari.515117, it was noticed that a small plastic tip protrudes from the puncture spike, which gets into the bottle when puncturing.As a result, about 100 protectors were opened, of which a good 30 are affected.The batch has now been put to one side.It seems that the puncture spike was not designed properly during production.".
 
Additional Manufacturer Narrative
The following fields were updated due to additional information: d10: device available for eval yes.D10: returned to manufacturer on: 28-dec-2022.H6: investigation summary several samples were provided to our quality team for investigation.Through visual inspection, the tip of the protector spike is observed to be bent in all of the samples.The shank of the protector was inspected and no damage or other issues were identified.Three retained samples were used for additional evaluation.All product was inspected and no damage was observed on any of the protector spikes.A device history review was performed for the reported lot 2201140, no deviations or non-conformances were identified during the manufacturing process that could have contributed to this issue.Product undergoes inspections throughout the manufacturing process to ensure the quality and functionality of the device, no issues were identified during manufacturing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD PHASEAL¿ PROTECTOR (P55)
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
BECTON DICKINSON, S.A.
camino de valdeolivia
s/n
san agustin de guadalix
Manufacturer (Section G)
BECTON DICKINSON, S.A.
camino de valdeolivia
s/n
san agustin de guadalix
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15544476
Report Number3003152976-2022-00455
Device Sequence Number6511790
Product Code ONB
UDI-Device Identifier00382905151170
UDI-Public(01)00382905151170
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K123213
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number515117
Device Lot Number2201140
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/13/2022
Supplement Date Received by Manufacturer01/11/2023
Initial Report FDA Received Date10/05/2022
Supplement Report FDA Received Date01/30/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/04/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-