Model Number 15
Medical Device Problem Code Failure to Power Up (1476)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/14/2022
Type of Reportable Event Malfunction
Event or Problem Description
The customer contacted stryker to report that their device would not power on with ac or dc power.In this state the device would be inoperable and defibrillation therapy would not be available if needed.There was no patient involvement reported with the event.
 
Additional Manufacturer Narrative
Stryker evaluated the customer's device and was able to verify and duplicate the reported issue.The power pcba was replaced to resolve the reported issue.After completing other unrelated repairs, proper device operation was observed through functional and performance testing.The device was returned to the customer for use.
 
Additional Manufacturer Narrative
Stryker further evaluated the removed power pcba and determined the cause of the reported issue was due to a shorted diode package cr20.
 
Event or Problem Description
The customer contacted stryker to report that their device would not power on with ac or dc power.In this state the device would be inoperable and defibrillation therapy would not be available if needed.There was no patient involvement reported with the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK® 15 DEFIBRILLATOR/MONITOR
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
todd bandy
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key15544554
Report Number0003015876-2022-02216
Device Sequence Number15099462
Product Code MKJ
UDI-Device Identifier00883873871638
UDI-Public00883873871638
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K142430
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 01/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number15
Device Catalogue Number99577-001963
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/21/2022
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/14/2022
Supplement Date Received by Manufacturer01/06/2023
Initial Report FDA Received Date10/05/2022
Supplement Report FDA Received Date01/11/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/25/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-