• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-0800-53
Device Problems False Positive Result (1227); Device Displays Incorrect Message (2591)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/23/2022
Event Type  malfunction  
Event Description
It was reported that during use on a patient, the cardiosave intra-aortic balloon pump (iabp) had a repeated low helium alarm.
 
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Manufacturer Narrative
Getinge field service engineer evaluated the unit, but was unable to reproduce the reported malfunction.The fse replaced the helium regulator as a preventive measure.The fse exercised the unit with no fail.The fse performed a full calibration and functional check per the factory calibration procedures.The iabp was returned to the customer and cleared for clinical use.The following investigation was performed the technician of the maquet failure analysis and testing dept.(fat) wayne, nj.The helium regulator was observed per the cardiosave service manual with visual damage.The failure analysis and testing dept.Could not install the helium regulator into the cardiosave test fixture and verify the leak as per factory specifications per procedure and the cardiosave service manual.Helium transducer regulator could not be tested due to physical damage.Helium transducer regulator will be held in the fat dept.Per procedure.
 
Event Description
It was reported that during use on a patient, the cardiosave intra-aortic balloon pump (iabp) had a repeated low helium alarm.There was no patient harm reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID, TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key15544593
MDR Text Key302831925
Report Number2249723-2022-02545
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108391
UDI-Public10607567108391
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 06/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/05/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-0800-53
Device Catalogue Number0998-00-0800-53
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/07/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/22/2012
Is the Device Single Use? No
Type of Device Usage Reuse
Removal/Correction Number2249723-01/24/2023-003-C
Patient Sequence Number1
-
-