BOSTON SCIENTIFIC CORPORATION CHOICE PT; WIRE, GUIDE, CATHETER Back to Search Results
Model Number 2080
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event: date of event: used the first date of the month of the aware date as no event date was provided.
 
Event or Problem Description
It was reported that guide wire tip detachment occurred.A 300 pt es j choice pt guidewire was advanced for treatment.However, during the procedure, it was noted that the tip area of the guidewire was fractured.The procedure was completed with a different device.There were no patient complications reported.
 
Additional Manufacturer Narrative
B3 - date of event: date of event: used the first date of the month of the aware date as no event date was provided.Device evaluated by mfr.: the device was returned for analysis.Unit returned identified distal tip kinked.The device was measured in three sections (distal - medium - proximal) along it in order to confirm that is within specification.Boston scientific manufacturing processes include extensive inspections to ensure that all finished devices meet specifications.
 
Event or Problem Description
It was reported that guide wire tip detachment occurred.A 300 pt es j choice pt guidewire was advanced for treatment.However, during the procedure, it was noted that the tip area of the guidewire was fractured.The procedure was completed with a different device.There were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHOICE PT
Common Device Name
WIRE, GUIDE, CATHETER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
302 parkway, global park
la aurora - heredia
CS  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15544967
Report Number2124215-2022-38493
Device Sequence Number5684598
Product Code DQX
UDI-Device Identifier08714729176947
UDI-Public08714729176947
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K143587
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/15/2024
Device Model Number2080
Device Catalogue Number2080
Device Lot Number0028687249
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/21/2022
Supplement Date Received by Manufacturer11/22/2022
Initial Report FDA Received Date10/05/2022
Supplement Report FDA Received Date11/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/15/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
Patient RaceAsian
-
-