MENTOR TEXAS MENTOR MEMORYSHAPE BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Model Number 3341152
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Hematoma (1884); Post Operative Wound Infection (2446)
Date of Event 08/01/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The product was returned to mentor for evaluation.Mentor conducted a visual inspection of the returned device.Device evaluation: during visual evaluation no apparent damage or visual anomalies were observed on the returned device.Infection, manifested by swelling, tenderness, pain and fever, may appear in the immediate postoperative period or at any time after insertion of the tissue expander.In the absence of classic symptoms, subacute or chronic infections may be difficult to diagnose.If infection does not subside promptly with the appropriate treatment, removal of the tissue expander is indicated.Infection is a known complication associated with any surgery.Per a database query, this is the only reported complaint of infection against this lot number.The mentor microbiology department has provided information on the sterility lot for the implanted device indicating that it met all sterilization parameters required to provide a 10e-6 sterility assurance level prior to release for distribution.Product evaluation concluded the reported infection occurred from a source other than the device.Infection is a known complication associated with any surgery and is referenced in our current product insert data sheet.Hematoma is a collection of blood within the space around the implant.Symptoms from hematoma may include swelling, pain, and bruising.While the body absorbs small hematomas, some will require surgery.If a hematoma occurs, it will most likely occur soon after surgery; however, it can occur at anytime.Hematoma is a known complication associated with this procedure and is referenced in our product insert data sheet.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Reason for device explant and/or reoperation: hematoma, wound infection.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a 73-year-old caucasian female patient underwent a primary breast augmentation surgery with a 345cc mentor memoryshape breast implant.Post-operatively, the patient experienced a wound infection and a hematoma on the left side, which was confirmed during a physical exam.As a result, the patient underwent removal on (b)(6) 2022.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR MEMORYSHAPE BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
michelleann garcia
3041 skyway circle north
irving, TX 75038
646591-7981
MDR Report Key15545000
Report Number1645337-2022-11742
Device Sequence Number12389815
Product Code FTR
UDI-Device Identifier00081317005193
UDI-Public00081317005193
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P060028
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number3341152
Device Catalogue Number3341152
Device Lot Number9725064
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/28/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/13/2022
Initial Report FDA Received Date10/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/22/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age73 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-