The product was returned to mentor for evaluation.Mentor conducted a visual inspection of the returned device.Device evaluation: during visual evaluation no apparent damage or visual anomalies were observed on the returned device.Infection, manifested by swelling, tenderness, pain and fever, may appear in the immediate postoperative period or at any time after insertion of the tissue expander.In the absence of classic symptoms, subacute or chronic infections may be difficult to diagnose.If infection does not subside promptly with the appropriate treatment, removal of the tissue expander is indicated.Infection is a known complication associated with any surgery.Per a database query, this is the only reported complaint of infection against this lot number.The mentor microbiology department has provided information on the sterility lot for the implanted device indicating that it met all sterilization parameters required to provide a 10e-6 sterility assurance level prior to release for distribution.Product evaluation concluded the reported infection occurred from a source other than the device.Infection is a known complication associated with any surgery and is referenced in our current product insert data sheet.Hematoma is a collection of blood within the space around the implant.Symptoms from hematoma may include swelling, pain, and bruising.While the body absorbs small hematomas, some will require surgery.If a hematoma occurs, it will most likely occur soon after surgery; however, it can occur at anytime.Hematoma is a known complication associated with this procedure and is referenced in our product insert data sheet.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Reason for device explant and/or reoperation: hematoma, wound infection.Manufacturer¿s reference number: (b)(4).
|
|
It was reported that a 73-year-old caucasian female patient underwent a primary breast augmentation surgery with a 345cc mentor memoryshape breast implant.Post-operatively, the patient experienced a wound infection and a hematoma on the left side, which was confirmed during a physical exam.As a result, the patient underwent removal on (b)(6) 2022.
|