MAUDE Adverse Event Report: EXACTECH, INC. CR TIBIAL INSERT SZ 2, 11MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. CR TIBIAL INSERT SZ 2, 11MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Model Number 200-22-11
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 06/21/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Pending evaluation.Concomitant device(s): 2168310 200-02-32 - three peg patella 32mm, 2246099 230-03-02 - optetrak asy,cr cemented femoral, sz 2, 2260478 02-012-45-2020 - lgc tibial fit tray cem sz 2f / 2t.
 
Event or Problem Description
It was reported that this female patient's right knee was revised 10 years post op due to premature polyethylene wear of the tibial insert.Patient had loose components and osteolysis.Following the revision surgery, plaintiff continues to be limited in her activities of daily living.Despite undergoing revision surgery, plaintiff experiences daily pain and discomfort in her right knee which limits her activities of daily living and impacts her quality of life.No other information available.
 
Additional Manufacturer Narrative
Section h10: (h3) based on review of all available information, there is no evidence to suggest that the reported event is related to any design issues.The cause of the subsequent revision cannot be conclusively determined; however, it may have been the result of malalignment between the femoral and tibial components, third body wear, patient-related conditions, or any combination of these possibilities, which led to prosthesis wear.Section h11: *the following sections have corrected information: (b5) describe event or problem: it was reported via legal that this female patient's right knee was revised 10 years post op due to polyethylene wear of the tibial insert; patient had loose components and osteolysis.No other information available.
 
Event or Problem Description
It was reported via legal that this female patient's right knee was revised 10 years post op due to polyethylene wear of the tibial insert; patient had loose components and osteolysis.No other information available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CR TIBIAL INSERT SZ 2, 11MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 ct
gaiensville FL 32653
Manufacturer Contact
kate jacobson
3523771140
MDR Report Key15545512
Report Number1038671-2022-01248
Device Sequence Number5685128
Product Code JWH
UDI-Device Identifier10885862040206
UDI-Public10885862040206
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K932776
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Consumer,Health Professional,Company Representative
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 11/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/11/2016
Device Model Number200-22-11
Device Catalogue Number200-22-11
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/15/2022
Supplement Date Received by Manufacturer10/25/2022
Initial Report FDA Received Date10/05/2022
Supplement Report FDA Received Date11/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/12/2008
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0019-2022
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-