| Model Number |
021299-10 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
09/02/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the single-use sterile valve seal was inserted into a laparoscopic trocar & cannula port as a surgical instrument for laparoscopic surgery.The scrub nurses found the valve seal was broken with a missing part after surgeons removed the cannula and returned it to the scrub nurses on the table.As the broken piece of the valve is missing.Internal reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The affected device has been requested for investigation by the manufacturer.
|
| |
|
Additional Manufacturer Narrative
|
|
There have been returned 3 unused valve seals (sterile packed) of the same charge as the failed.The devices in question got discarded by the user.These three devices got unpacked and inspected for damages and/or material degradation and tested by a simulated use with 50 passages of an instrument through the valves.The provided unused products of the same batch didn't show any defects - even after 50 simulated penetrations - no final cause for the described failure can be determined.The most probable root cause is that a damaged instrument with a sharp edge passed the valve and damaged them during use.Internal vigilance reference number for this case is (b)(4).
|
| |
|