• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KARL STORZ SE & CO. KG SEAL FOR LAPAROSCOPE TROCAR SLEEVE; VALVE SEAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KARL STORZ SE & CO. KG SEAL FOR LAPAROSCOPE TROCAR SLEEVE; VALVE SEAL Back to Search Results
Model Number 021299-10
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 09/02/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the single-use sterile valve seal was inserted into a laparoscopic trocar & cannula port as a surgical instrument for laparoscopic surgery.The scrub nurses found the valve seal was broken with a missing part after surgeons removed the cannula and returned it to the scrub nurses on the table.As the broken piece of the valve is missing.Internal reference number: (b)(4).
 
Additional Manufacturer Narrative
The affected device has been requested for investigation by the manufacturer.
 
Additional Manufacturer Narrative
There have been returned 3 unused valve seals (sterile packed) of the same charge as the failed.The devices in question got discarded by the user.These three devices got unpacked and inspected for damages and/or material degradation and tested by a simulated use with 50 passages of an instrument through the valves.The provided unused products of the same batch didn't show any defects - even after 50 simulated penetrations - no final cause for the described failure can be determined.The most probable root cause is that a damaged instrument with a sharp edge passed the valve and damaged them during use.Internal vigilance reference number for this case is (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SEAL FOR LAPAROSCOPE TROCAR SLEEVE
Common Device Name
VALVE SEAL
Manufacturer (Section D)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM  78532
Manufacturer (Section G)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM   78532
Manufacturer Contact
anja fair
2151 e grand ave
el segundo, CA 90245-5017
4242188738
MDR Report Key15546133
Report Number9610617-2022-00286
Device Sequence Number6575807
Product Code GCJ
UDI-Device Identifier04048551400203
UDI-Public4048551400203
Combination Product (Y/N)N
Initial Reporter CountryHK
PMA/510(K) Number
K180977
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number021299-10
Device Catalogue Number021299-10
Device Lot Number17/48/203
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/19/2022
Supplement Date Received by Manufacturer10/26/2022
Initial Report FDA Received Date10/05/2022
Supplement Report FDA Received Date11/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-