EXACTECH, INC. NV GXL LNR, +5LAT, 32MM G1-48/50MM CUPS; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
|
Back to Search Results |
|
| Model Number |
NV GXL LNR, +5LAT, 32MM G1-48/50MM CUPS |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
09/07/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a 60 yo female patient, initial implant of the left knee on (b)(6) 2016, was revised due to early gxl poly wear on (b)(6) 2022, approximately 5 years 9 months post the initial implant date.They were revised to novation xle poly.Patient was last known to be in stable condition following the event.The devices are available for return.
|
| |
|
Additional Manufacturer Narrative
|
|
Concomitants: 4339937 188-00-05 - wedge plasma s/o sz 5, 4492908 142-32-00 - cocr fem head 32mm +0 offset 12/14, 4552056 180-01-50 - nv crown cup clstr hole 50mm group 1.(b)(4).Additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: based on the available information, the patient involved meets the following risk criteria for early prosthesis wear as specified; implanted with a lateralized liner, combination of largest available femoral head and/or thinnest available acetabular liner was used, and implanted with a component having a shelf age of greater than 2 years.The most likely cause for the revision reported due to prosthesis wear is a combination of the risk factors specified.
|
| |
|
Event or Problem Description
|
|
It was reported that a femalepatient, initial implant of the left knee and then was revised due to early gxl poly wear, approximately 5 years 9 months post the initial implant date.They were revised to novation xle poly.Patient was last known to be in stable condition following the event.The devices are available for return.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were added/corrected: b1, b2, b5, d1, g1, h6: clinical code, component code, investigation findings, investigation conclusions.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|