• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION PLATINUM PLUS; WIRE, GUIDE, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION PLATINUM PLUS; WIRE, GUIDE, CATHETER Back to Search Results
Model Number 39895
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/19/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that distal tip detachment occurred.The target lesion was located in the severely calcified femoral artery.A 018/260 platinum plus guidewire was used during the procedure.However, when it was retrieved from the patient, it was noticed that the floppy distal end was not attached to the guidewire.No further patient complications were reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: the device was not returned by the customer, nevertheless the customer attached a picture.It was observed that the coil wire was stretched.It is important to mention that the device is observed inside the pouch, nevertheless it is not possible to see the entire device.Boston scientific manufacturing processes include extensive inspections to ensure that all finished devices meet specifications.
 
Event or Problem Description
It was reported that distal tip detachment occurred.The target lesion was located in the severely calcified femoral artery.A 018/260 platinum plus guidewire was used during the procedure.However, when it was retrieved from the patient, it was noticed that the floppy distal end was not attached to the guidewire.No further patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PLATINUM PLUS
Common Device Name
WIRE, GUIDE, CATHETER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15546392
Report Number2124215-2022-37873
Device Sequence Number6575315
Product Code DQX
UDI-Device Identifier08714729035961
UDI-Public08714729035961
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K935997
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/08/2024
Device Model Number39895
Device Catalogue Number39895
Device Lot Number0027804512
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/19/2022
Supplement Date Received by Manufacturer10/20/2022
Initial Report FDA Received Date10/05/2022
Supplement Report FDA Received Date11/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/09/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-