Model Number NX029Z
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Code Bone Fracture(s) (1870)
Date of Event 09/09/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported to aesculap inc.That an as vega ps femoral comp.Cemented f4n r (part # nx029z) was implanted during a primary procedure performed in 2018.According to the complainant, there was postoperative loosening of the implant.This occurred due to a patient fall.The patient underwent a revision procedure on (b)(6) 2022.The complaint device has not been returned to the manufacturer for evaluation.A revision was required; columbus revision products were implanted.Further details have been requested.The adverse event / malfunction is filed under aic reference 400569176.Associated medwatch report: (b)(4) 2916714-2022-00101.
 
Additional Manufacturer Narrative
Manufacturing site evaluation: the device was not returned to the manufacturer for physical evaluation.Additionally, no on-site evaluation of the unit was performed and no parts were returned for failure analysis.Therefore, the investigation was not able to confirm a device issue that could be associated with the reported event.The device history records (dhr) were not able to be reviewed as no lot number was made available.However, all device history records (dhr) are reviewed and released according to documented procedures and a device is not released if it does not meet requirements or is nonconforming.A definitive conclusion regarding the complaint incident cannot be reached without a physical examination of the complaint device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AS VEGA PS FEMORAL COMP.CEMENTED F4N R
Common Device Name
KNEE ENDOPROSTHETICS
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
MDR Report Key15546409
Report Number2916714-2022-00100
Device Sequence Number12406085
Product Code JWH
UDI-Device Identifier04046963810122
UDI-Public4046963810122
Combination Product (Y/N)N
Initial Reporter StateNV
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Importer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 10/05/2022
Report Date (Section F) 10/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberNX029Z
Device Catalogue NumberNX029Z
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date09/09/2022
Event Location Hospital
Date Report to Manufacturer09/09/2022
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date10/05/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
Date Report Sent to FDA10/05/2022
-
-