| Model Number |
XPER FLEX CARDIO FC2010 REV C |
| Medical Device Problem Code |
Incorrect Measurement (1383)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/15/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Complaint evaluation: in an effort to resolve the reported issue remotely, philips technical support (ts) connected to the device and found the following; windows updates were performed on (b)(6) 2022.Service pack 2 for sql server 2014 ((b)(4)) - not on phillips approved list.Microsoft visual c 2010 x64 and x86 ((b)(4)) - not on phillips approved list.Microsoft visual shell (isolated) - enu ((b)(4)) - not on phillips approved list.Service pack 2 - microsoft vss writer for sql server 2014 ((b)(4)) - not on phillips approved list.Nic for flex was set to auto negotiate.Changed this to 100 full.Nic order needed to be moved up 1.Hospital was 10.Flex was 100 and set under hospital now.Firewall was on, turned this off.Uac was off.Uninstalled/reinstalled flex cardio.Flex versions match and are correct for a rev c.Device is set and registered as a host.Host shows it communicates with a nurse.Ip shown is (b)(6) (port 4023).However, users advised that they do not use any nurse device.Ts also could not connect or ping that ip address.Ts deleted this from the host settings.Device was recently rebooted.In addition, a philips field service engineer (fse) was dispatched to the customer's site to further evaluate the device.The device evaluation was performed by the fse on the actual device involved in this complaint.Customer resolution and conclusion: the absence of further calls supports that the reported problem was resolved.The device remains in use at the customer's facility.No further action or investigation is warranted based on the available information at the time of complaint closure.
|
| |
|
Event or Problem Description
|
|
The customer reported having intermittent monitoring issues in all rooms.According to the customer, users are experiencing intermittent issues with monitoring off error messages and the ecg is not matching the boom monitor.It was further stated by the customer that issues have been happening since their upgrade from software version 2.2 to 2.6 and when all devices were changed to windows 10.This case will address the ecg issue.The device did not behave as expected by the customer.To resolve the issue, the attending fse reimaged the device.The device was in use on a patient.There was no report of patient or user harm.
|
| |
|
Search Alerts/Recalls
|