• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION CRE PRO; BALLOON DILATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION CRE PRO; BALLOON DILATOR Back to Search Results
Model Number M00558660
Medical Device Problem Code Defective Device (2588)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/14/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported to boston scientific corporation that a cre pro wireguided dilatation balloon was used in the esophagus during a dilatation procedure performed on an (b)(6), 2022.During preparation, it was noticed that the balloon was bent at the top/balloon when taken out on the packaging.The procedure was completed with another cre pro wireguided dilatation balloon.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
Block h6: device code a0203 captures the reportable event of balloon bent.Block h10-: investigation results: the returned cre pro wireguided dilatation balloon was analyzed, and a visual examination found that the balloon and catheter of the device had no damages.Visual inspection of the device noted evidence that the wireguided was bent/kinked in the distal side.No other problems with the device were noted.With all the available information, boston scientific concludes the reported event was confirmed.The kink found in the wire are likely to have occurred due manipulation the device during the preparation.Forcing the device against significant resistance could result in device damage or loss of functionality the manner as the device was handled and manipulated may have caused the reported and encountered failure.This event is unlikely to cause or contribute to a death or unanticipated serious deterioration in health, nor would this event do so if it were to recur.Therefore, the most probable root cause is adverse event related to procedure.A labeling review was performed and, from the information available, this device was used per the directions for use (dfu) / product label.
 
Event or Problem Description
It was reported to boston scientific corporation that a cre pro wireguided dilatation balloon was used in the esophagus during a dilatation procedure performed on an (b)(6) 2022.During preparation, it was noticed that the balloon was bent at the top/balloon when taken out on the packaging.The procedure was completed with another cre pro wireguided dilatation balloon.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRE PRO
Common Device Name
BALLOON DILATOR
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
cork business technology park
model farm road
cork
EI  
Manufacturer Contact
carole morley morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key15547031
Report Number3005099803-2022-05721
Device Sequence Number1944283
Product Code FGE
UDI-Device Identifier08714729797548
UDI-Public08714729797548
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K112994
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/29/2024
Device Model NumberM00558660
Device Catalogue Number5866
Device Lot Number0028461770
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/14/2022
Supplement Date Received by Manufacturer01/05/2023
Initial Report FDA Received Date10/05/2022
Supplement Report FDA Received Date01/24/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/29/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-