| Model Number |
1001780J |
| Medical Device Problem Codes |
Entrapment of Device (1212); Material Separation (1562)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
08/22/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
User facility medwatch report received that states: mdr report #: mw5111748.During pci (percutaneous coronary intervention) for an acute mi (myocardial infarction), the end of the guidewire caught on the stent leaving a portion of the guidewire remaining caught on the stent.0.014" guidewire.Fda safety report id # (b)(4).Additionally, the following information was received: it was reported that the procedure was to treat a lesion located in the right coronary artery.After the guide wire separated, a snare was advanced to the separated portion of wire and a small piece was able to be snared and removed; however, another small piece of wire was left embedded behind the implanted stent.The patient was transferred to another hospital for second opinion related to this incident.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.Medwatch report #: mw5111748.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents/complaints from this lot.The investigation determined the reported difficulties and subsequent treatment appear to be related to the operational context of the procedure.There is no indication of a product quality issue with respect to manufacture, design, or labeling.D10, h3 - it was initially reported that the device would be returned for analysis.Subsequent information revealed that the device was discarded and is not available for evaluation.
|
| |
|
Search Alerts/Recalls
|