| Model Number |
640CX2535 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/29/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Manufacturer¿s ref.No: (b)(4).Information regarding patient identifier, date of birth, age, gender, weight, race, and ethnicity were not provided.Procode is krd/hcg.The name, phone and email address of the initial reporter are not available / reported.The device is available to be returned for evaluation and testing.However, it has not been received to date as indicated as ¿other¿ in this section as the reason for non-evaluation.If the device returns, a device investigation will be performed.A review of manufacturing documentation associated with this lot (30739311) presented no issues during the manufacturing or inspection processes related to the reported complaint.There were no nonconformances related to device manufacture or inspection.All product rejected during manufacturing was identified as scrap and properly accounted for.This report is being submitted pursuant to the provisions of 21 cfr, part 4.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by cerenovus, or its employees that the report constitutes an admission that the product, cerenovus, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
|
| |
|
Event or Problem Description
|
|
The healthcare professional reported that during an endovascular embolization procedure, the 4mm x 8cm orbit galaxy complex xtrasoft coil (640cx0408 / 30753131) was the first coil used.The physician decided to pull it out for re-deployment after trying another size of coil, because the coil was not of an appropriate size.During the attempt to re-sheath the coil, the introducer failed to be re-zipped.The physician used another ¿same like¿ product as a replacement.A 2.5mm x 3.5cm orbit galaxy complex xtrasoft coil (640cx2535 / 30739311) was also deployed as the third coil.However, when the physician tried to detach the coil, it could not be detached; the physician thought the hub appeared broken so there was not enough pressure delivered to the detachment zone.The physician observed saline leaking from the hub.The physician replaced this coil with another 2.5mm x 3.5cm orbit galaxy complex xtrasoft coil and the replacement coil successfully detached and the procedure was successfully completed.There was no report of any negative impact to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturer¿s ref.No: (b)(4).The purpose of this mdr submission is to report that the product was received in the product analysis lab on 20-oct-2022.The returned product is pending evaluation.A supplemental 3500a report will be submitted once the product investigation has been completed.Updated sections: b.4, d.9, g.3, g.6.H.2, h.3, h.6, and h.10.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturer¿s ref.No: (b)(4).The purpose of this mdr submission is to report the investigational finding of the returned device.[conclusion]: the healthcare professional reported that during an endovascular embolization procedure, the 4mm x 8cm orbit galaxy complex xtrasoft coil (640cx0408 / 30753131) was the first coil used.The physician decided to pull it out for re-deployment after trying another size of coil, because the coil was not of an appropriate size.During the attempt to re-sheath the coil, the introducer failed to be re-zipped.The physician used another ¿same like¿ product as a replacement.A 2.5mm x 3.5cm orbit galaxy complex xtrasoft coil (640cx2535 / 30739311) was also deployed as the third coil.However, when the physician tried to detach the coil, it could not be detached; the physician thought the hub appeared broken so there was not enough pressure delivered to the detachment zone.The physician observed saline leaking from the hub.The physician replaced this coil with another 2.5mm x 3.5cm orbit galaxy complex xtrasoft coil and the replacement coil successfully detached and the procedure was successfully completed.There was no report of any negative impact to the patient.Multiple attempts to obtain additional information related to the procedure and the reported device issue were unsuccessful.If additional information is received at a later date, this complaint file will be updated accordingly.The complaint device was returned for evaluation and analysis.The investigation finding is documented below.Investigation summary: a non-sterile 2.5mm x 3.5cm orbit galaxy complex xtrasoft coil was received contained in the decontamination pouch.Visual inspection was performed.It was observed that the embolic coil was outside of the introducer.The hub was visually inspected and no appearance of damages were observed on the hub.The returned device underwent microscopic inspection.Under magnification, it was noted that the embolic coil is stretched and has multiple kinked sections.Some residues of dried blood and dried saline solution were observed on the embolic coil.No abnormalities were observed on the hub under magnification.Coil stretching and coil kinking are known potential issues associated with the use of this device.The instructions for use (ifu) provides proper handling instructions for the device to prevent such issues.Stretching can occur during procedure handling where force may have been inadvertently applied.The coil kinking and coil stretching were not originally reported in the compliant and may have occurred during the coil replacement.Such coil damages are not related to the hub leakage and subsequent failure to detach the coil.The complaint device was connected to a lab sample syringe.It was then flushed until the gauge reached the green zone; and after approximately 3 seconds, the coil was successfully detached from the unit.No leak was observed at the hub during this step.Based on this evidence, the reported issue documented in the complaint regarding a luer hub leakage could not be confirmed.It is possible that when the healthcare professional fitted the hub of the delivery tube to the luer valve, this was not properly engaged, which resulted in the leakage (insufficient pressure) observed during the procedure; however, this cannot be conclusively determined.No damages were found in the device that may have contributed to the issue experienced during the procedure.There is no indication that the issue reported in the complaint results from a defect inherently related to the device.A review of manufacturing documentation associated with this lot (30739311) presented no issues during the manufacturing or inspection processes related to the reported complaint.There were no nonconformances related to device manufacture or inspection.As part of the cerenovus quality process, all devices are manufactured, inspected, and released to approved specifications.It should be noted that multiple factors could cause failure.The instructions for use (ifu) does contain the following recommendation: if saline fluid leaks from the syringe, or the threaded plunger strips and does not allow the syringe to obtain the required pressure anytime during the detachment procedure, the syringe should be replaced.Based on the manufacturing documentation review, there is no indication that the event is related to the device manufacturing process.As part of the post market surveillance program, information from this complaint is trended for statistical signals and corrective / preventive action may be triggered at a later time.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.H.6: the code ¿no device problem found (c19)¿ was used in the investigation findings because the reported issue complaint regarding a luer hub leakage could not be confirmed based on the functional evaluation performed on the returned complaint device.This code corresponds with the ¿no problem detected (d14)¿ in the investigation conclusion.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
|
| |
|