Model Number 4018PI
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 09/23/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Healthcare professional called in and stated they had placed a port for a cancer patient, and noticed after a year that there is some leaking.They removed the port and replaced it.The patient condition was reported as fine post-procedure.
 
Additional Manufacturer Narrative
The potential lot affected is lot number 18050137.The review of the manufacturing batch record of the polysite 4018pi implantable port final product (lot 18050137) did not highlight any deviation that could be the root cause of the incident.Indeed, the sealing of the port-catheter-connection ring assembly is verified on 100% of the devices manufactured by injecting air at a pressure of 2.2 bar +/- 0.2 bar for 1 minute.Alteration of the catheter or connecting ring would most likely have led to a leakage and would have been detected.The leak test performed on the (b)(4) units manufactured is in accordance with the specifications.No deviation was detected.The device is in transit for returned to perouse medical.Perouse medical hasn't been informed of other incidents implying the batch 18050137 (potential lot).We asked more information about the incident (date implantation, date explantation, venous access routes, implantation and explantation reports, copy of x-ray pictures for implantation and observation of event, the consequences for the patient).We are waiting for the device in order to perform a technical investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLYSITE PI IMPLANTABLE INFUSION PORT
Common Device Name
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer Contact
agnès jordan
4 rue saint-anne
ivry-le-temple, 60173
FR   60173
MDR Report Key15548087
Report Number3006942524-2022-02358
Device Sequence Number1905240
Product Code LJT
UDI-Device Identifier03661234017638
UDI-Public03661234017638
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K122834
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 10/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Model Number4018PI
Device Lot Number18050137
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/26/2022
Initial Report FDA Received Date10/06/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-