• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN MEDICAL INC. PERIFIX®; EPIDURAL ANESTHESIA KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN MEDICAL INC. PERIFIX®; EPIDURAL ANESTHESIA KIT Back to Search Results
Model Number 332215
Medical Device Problem Code Break (1069)
Health Effect - Clinical Codes Inadequate Pain Relief (2388); Foreign Body In Patient (2687)
Date of Event 08/24/2022
Type of Reportable Event Malfunction
Event or Problem Description
Per medwatch number mw5111787: description: epidural catheter placed in laboring mom.Patient had inadequate pain relief.When removed from patient to attempt another catheter, it was identified the tip of the catheter was missing.
 
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number 400569838.The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
 
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).No sample was provided for evaluation.Based on the data from the investigation we are unable to determine the root cause of the reported incident.The reported defect was unable to be confirmed.The actual defective device is a valuable tool in investigating the cause of this incident.Review of the discrepancy management system (dsms) database was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERIFIX®
Common Device Name
EPIDURAL ANESTHESIA KIT
Manufacturer (Section D)
B. BRAUN MEDICAL INC.
901 marcon blvd.
allentown PA 18109
Manufacturer Contact
jonathan severino
901 marcon blvd.
allentown, PA 18109
4847197287
MDR Report Key15549637
Report Number2523676-2022-00498
Device Sequence Number1919313
Product Code OGE
UDI-Device Identifier04046964178245
UDI-Public(01)04046964178245
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2023
Device Model Number332215
Device Catalogue Number332215
Device Lot Number0061800921
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/19/2022
Supplement Date Received by Manufacturer09/19/2022
Initial Report FDA Received Date10/06/2022
Supplement Report FDA Received Date11/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/04/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age22 YR
Patient SexFemale
Patient Weight99 KG
-
-