Model Number 21303875
Medical Device Problem Code Biocompatibility (2886)
Health Effect - Clinical Codes Hypersensitivity/Allergic reaction (1907); Inflammation (1932); Local Reaction (2035); Ear infection (4813); Dry Skin (4839)
Date of Event 07/01/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Manufacturer's ref# (b)(4).Technical investigation concluded: a device history record review have been completed.No deviation or changes were found during manufacturing of the device.The clinical investigation concluded: allergic reaction is identified in the risk analysis and mitigated to an acceptable level through device design by compliance with standards for biocompatibility.Still allergic reactions can occur in rare cases.The clinical evaluation does evaluate allergic reactions and the user guide includes a safety notification instructing the hearing aid wearer to contact the hcp in case of skin irritation.There are no new clinical aspects (e.G.Unforeseen use or clinical conditions) discovered from this event.Therefore, it is concluded that the current clinical evaluation and risk/benefit conclusions herein are sufficient.Risk register reference: the residual risk after the risk control measure(s) is acceptable.No new hazardous situations revealed and no need to update the risk register based on the above information the case is considered reportable as the event / incident did lead to, or might led to, death or a serious injury.This is a final report.
 
Event or Problem Description
After being fit with the resound one 7 rie devices, clinician reported that end user received a severe reaction to the new light grey domes, resulting in inflammation and bleeding within both ear canals.Portion in front of the ear (anterior to tragus and helix) was also reported to have some irritation from the receivers.The reaction began a couple of days after initial fitting, and the end user could not use the device until follow up appointment, which was two weeks after.No known history of allergies or hypersensitivity.The user was in contact with authorized medical practitioner, who reported it to be an allergic reaction of some kind to the devices.Topical cream was prescribed for treatment.Recommendation from health care professional was to fit custom moulds with hypoallergenic coating.Clinician reported that the user's ear has since healed and are now wearing canal lock moulds with no reported allergic reactions.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GN RESOUND
Common Device Name
HEARING AID, AIR CONDUCTION WITH WIRELESS TECHNOLOGY
Manufacturer (Section D)
GN HEARING A/S
lautrupbjerg 7
ballerup, 2750
DA  2750
Manufacturer Contact
karen frisk
lautrupbjerg 7
ballerup, 2750
DA   2750
MDR Report Key15549673
Report Number3005650109-2022-00045
Device Sequence Number19312527
Product Code OSM
UDI-Device Identifier05708296194575
UDI-Public05708296194575
Combination Product (Y/N)N
Initial Reporter CountryAS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 10/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number21303875
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/07/2022
Initial Report FDA Received Date10/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/26/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age88 YR
Patient SexMale
-
-