| Model Number |
321-25-42 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/15/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
As reported, during an initial tsa surgery, when using the 42 reamer, the pilot tip snapped off within the glenoid and had to be retrieved from the bone using a bone nibbler.The patient was last known to be in stable condition following the event.No other patient information/medical history reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Pending device return and investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
The broken device reported was likely the result of applying a bending moment to the reamer during use, which led to brittle fracture of the pilot tip feature.
|
| |
|