DEPUY SYNTHES PRODUCTS LLC SAGITTAL MICRO SAW; MOTOR, SURGICAL INSTRUMENT, AC-POWERED Back to Search Results
Model Number S-SAW
Medical Device Problem Codes Excessive Heating (4030); Difficult or Delayed Separation (4044)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.Additional narrative: device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the device generating heat identified during service and evaluation was confirmed.The assignable root cause was determined to be due to component failure from wear.Udi: (b)(4).
 
Event or Problem Description
It was reported by japan that during service and evaluation, it was determined that the sagittal microsaw attachment device generated heat and experienced vibration.It was further observed that the device was difficult to assemble/disassemble, was stuck in connection, and made an unexpected noise.It was further determined that the device failed pretest for attachment engagement assessment and temperature assessment.It was noted in the service order that the device was difficult to connect and remove.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2022.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: h6: device investigation update: upon further investigation of the device, it was further determined that the cutter device could not be removed.H6.Investigation findings and component codes have been updated accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAGITTAL MICRO SAW
Common Device Name
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
kate karberg
4500 riverside drive
palm beach gardens, FL 33410
3035526892
MDR Report Key15550293
Report Number1045834-2022-01348
Device Sequence Number17661550
Product Code GEY
UDI-Device Identifier00845384008521
UDI-Public00845384008521
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K080802
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberS-SAW
Device Catalogue NumberS-SAW
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/29/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/04/2022
Supplement Date Received by Manufacturer10/12/2022
Initial Report FDA Received Date10/06/2022
Supplement Report FDA Received Date10/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/14/2014
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-