| Model Number |
1570-02-135 |
| Medical Device Problem Codes |
Degraded (1153); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Foreign Body Reaction (1868); Hypersensitivity/Allergic reaction (1907); Pain (1994); Loss of Range of Motion (2032); Metal Related Pathology (4530); Physical Asymmetry (4573); Swelling/ Edema (4577)
|
| Date of Event |
03/28/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).The device catalog number is unknown; therefore, udi is unavailable.Occupation: lawyer.Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Pinnacle litigation complaint received ad 20 sep 2022.Patient was revised was due to pain, toxic levels of cobalt and chromium, mri showed consistent metallosis.Doi: 2011, dor: (b)(6) 2022, right hip.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.A worldwide lot specific complaint database search, or manufacturing record evaluation, was not possible as the required lot number was not provided.The information received will be retained for trend analysis, post market surveillance, or other events within the quality system.Device history lot : a manufacturing record evaluation (mre), was not possible because the required lot code was not provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint (b)(4).Investigation summary; the device associated with this report was not returned to depuy synthes for evaluation.X-ray evidence provided was reviewed and no visible evidence of a device nonconformance was found during the investigation.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot; a manufacturing record evaluation (mre), was not possible because the required lot code was not provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
After a review of the medical records, the event on 8/31/2022 reported closed posterior hip dislocation due to a waterskiing accident.The patient complained of right hip pain and limited movement.The patient had an obvious deformity of the right hip.The right leg was shortened and showed signs of hip dislocation.The surgeon was able to relocate the right hip.On (b)(6) 2022 patient fall from a standing position and heard an audible pop sound and had right hip dislocation, pain, tenderness, and distress.Recurrent dislocation and nerve damage.Dislocation reduction was performed.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.Added: a2 (age), b1(product problem), b5, b7, h6 health effect - clinical code.H6: swelling/ edema (e2338) is being utilized to capture swelling & edema.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Medical records received.After review of medical records patient was revised was due to failed right hip metal on metal replacement and elevated metal ions associated with pseudotumor.Operative notes indicated swelling prior to incision.Once the fascia was incised, a large serosanguineous that was great tented collection fluid was encountered.It was quite voluminous.That fluid collection was evacuated with suction.There was a large membrane of a great into amount of tissue coming from the joint and going into the extra-articular and lateral trochanteric tissues.The pseudotumor was at least 15x 20cm in size.N.There is significant corrosion seen at the base of the trunnion as well as within the ball.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).Investigation summary: the device associated with this report was not returned to depuy synthes for evaluation.The photographs attached were reviewed, however they do not represent the current complaint condition.Therefore the investigation could not draw any conclusions about the reported event.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: a manufacturing record evaluation was performed for the finished device (157002135/ek8fba) product and lot numbers, and no non-conformances were identified.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary the device associated with this report was not returned to depuy synthes for evaluation.The photographs attached were reviewed, however they do not represent the current complaint condition.Therefore the investigation could not draw any conclusions about the reported event.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot a manufacturing record evaluation was performed for the finished device ((b)(4).) product and lot numbers, and no non-conformances were identified.
|
| |
|
Search Alerts/Recalls
|