Model Number PLC141000
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Aneurysm (1708)
Date of Event 08/25/2021
Type of Reportable Event Serious Injury
Event or Problem Description
On (b)(6) 2021, the patient underwent endovascular treatment of a thoracoabdominal aortic aneurysm using gore® excluder® thoracoabdominal branch endoprostheses (tambe) and gore® excluder® aaa contralateral leg endoprostheses and gore® excluder® iliac branch endoprosthesis in the aaa1701 tambe study.On (b)(6) 2021, the patient was discharged.It was reported that images dated (b)(6) 2021, revealed a dissection of left external artery.
 
Additional Manufacturer Narrative
Patient medications: acetaminophen, aspirin, and oxycodone.The gore® excluder® aaa endoprosthesis instructions for use (ifu) states that adverse events that may occur and/or require intervention include, but are not limited to vascular trauma (e.G., ilio-femoral vessel dissection).W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Additional Manufacturer Narrative
Email from gore clinical study manager, the physician answered the questions: is the dissection (adverse event #5) in the external iliac artery originating immediately adjacent to the proximal or distal extents of the gore® excluder® aaa contralateral leg endoprosthesis: no.Is the dissection (adverse event #5) in the external iliac artery originating immediately adjacent to the proximal or distal extents of the gore® excluder® aaa iliac branch component: no.Was a previous dissection in the external iliac artery identified prior to the procedure: no.Was there a reintervention to correct the dissection: no.If yes on reintervention, what was the date and correction: n/a.Is the root cause of the external iliac artery dissection known: procedural.Additional investigation determined no product problem or deficiency caused or contributed to an adverse event/incident for the patient; the initial medwatch and any supplemental report submitted under manufacturer report number 3013164176-2022-01496 was submitted in error, and are hereby retracted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® EXCLUDER® AAA ENDOPROSTHESIS
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL SILICON VALLEY B/P
2890 de la cruz blvd.
santa clara CA 95050
Manufacturer Contact
rachael chascsa
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key15551058
Report Number3013164176-2022-01496
Device Sequence Number5959038
Product Code MIH
UDI-Device Identifier00733132618637
UDI-Public00733132618637
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
P020004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Reporter Type Manufacturer
Report Source Study,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/27/2024
Device Model NumberPLC141000
Device Catalogue NumberPLC141000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/15/2022
Supplement Date Received by Manufacturer09/15/2022
Initial Report FDA Received Date10/06/2022
Supplement Report FDA Received Date03/29/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/29/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient Age70 YR
Patient SexFemale
Patient Weight65 KG
-
-