MAUDE Adverse Event Report: STERIS MEXICO, S. DE R.L. DE C.V. AMSCO 400 STERILIZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STERIS MEXICO, S. DE R.L. DE C.V. AMSCO 400 STERILIZER Back to Search Results
Model Number AMSCO 400 STEAM STERILIZER
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Burn(s) (1757)
Date of Event 09/07/2022
Type of Reportable Event Malfunction
Event or Problem Description
The user facility reported that an employee was adjusting a rack within their amsco 400 sterilizer and the guide that holds the rack in place detached subsequently causing the employee to obtain a burn to their elbow.It is unknown if the employee sought or received medical treatment.
 
Additional Manufacturer Narrative
A steris service technician inspected the amsco 400 sterilizer and found screws that hold the metal guide plate inside the sterilizer chamber to be missing.The unit has been removed from service and is pending repairs.A follow-up report will be submitted when additional information becomes available.
 
Additional Manufacturer Narrative
A steris service technician made repairs to the sterilizer, tested the unit and confirmed it to be operating properly.The technician returned the sterilizer to service and no additional issues have been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMSCO 400 STERILIZER
Common Device Name
STERILIZER
Manufacturer (Section D)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, 67190
MX  67190
Manufacturer (Section G)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, 67190
MX   67190
Manufacturer Contact
daniel davy
5960 heisley road
mentor, OH 44060
4403927453
MDR Report Key15551100
Report Number3005899764-2022-00050
Device Sequence Number6077047
Product Code FLE
UDI-Device Identifier00724995209629
UDI-Public00724995209629
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberAMSCO 400 STEAM STERILIZER
Device Catalogue NumberSR0202210321
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/07/2022
Supplement Date Received by Manufacturer09/07/2022
Initial Report FDA Received Date10/06/2022
Supplement Report FDA Received Date11/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/16/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-