• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ENCORE 26 ADVANTAGE KIT; SYRINGE, BALLOON INFLATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ENCORE 26 ADVANTAGE KIT; SYRINGE, BALLOON INFLATION Back to Search Results
Model Number 2760
Medical Device Problem Code Inflation Problem (1310)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/29/2021
Type of Reportable Event Malfunction
Event or Problem Description
Reportable based on device analysis completed on 14sep2022.It was reported that the pressure lost, and manometer was dysfunctional.An encore 26 advantage kit inflation device was selected for use.During the procedure, it was noted that the pressure was lost, and the manometer was dysfunctional.The procedure was completed, and no patient complications were reported.However, device analysis revealed that the gauge accuracy test was not able to complete due to loss of the needle to zero location.
 
Additional Manufacturer Narrative
Device evaluated by manufacturer: the encore ic was returned for analysis.A visual inspection was conducted.The gauge needle was at 20 atm when received.Functional inspection was conducted, the needle can move when increasing air pressure using the plunger and when releasing it gets back to 20 atm.A pressure damping test was conducted and the encore ic passed the test.A vacuum test was conducted, there was no bubble leakage observed.The encore ic passed side load testing.The encore ic gauge accuracy test was not able to be completed due to the loss of the needle zero location.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENCORE 26 ADVANTAGE KIT
Common Device Name
SYRINGE, BALLOON INFLATION
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
model farm road
cork
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15551286
Report Number2134265-2022-08846
Device Sequence Number19538408
Product Code MAV
UDI-Device Identifier08714729180005
UDI-Public08714729180005
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K140745
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/14/2023
Device Model Number2760
Device Catalogue Number2760
Device Lot Number0028200298
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/24/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/14/2022
Initial Report FDA Received Date10/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/14/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-