• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM Back to Search Results
Model Number 1804250-33
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Perforation of Vessels (2135)
Date of Event 09/30/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.The reported patient effect of perforation is listed in the xience skypoint everolimus eluting coronary stent systems instructions for use as a known patient effect of coronary stenting procedures.A conclusive cause for the reported patient effect, and the relationship to the product, if any, cannot be determined; however, the subsequent treatment appears to be related to the operational context of the procedure.There is no indication of a product quality issue with respect to manufacture, design or labeling.The nc trek balloon referenced in is being filed under a separate medwatch report number.
 
Event or Problem Description
It was reported that the patient presented with pain and the procedure was to treat the left anterior descending (lad) artery and the left main (lm) artery with heavy calcification and mild tortuosity.Two xience skypoints were implanted in the lad and lm arteries.The nc trek balloon dilatation catheter (bdc) had no resistance during advancement and was used for post dilatation per standard procedure, however, a perforation occurred in the lad.An additional non-abbott stent was used to treat the perforation.The balloon was used again for post-dilatation and was fully deflated (confirmed by x-ray).However, the balloon would not come out of the guide catheter.Force was applied.All devices were removed together as a unit, however, the balloon separated from the shaft and got stuck in the lm artery.Intervention was attempted to embed the balloon but could not get a guidewire past the balloon.The patient died.Intravascular ultrasound (ivus) was used and there was no thrombus.In the physician's opinion, the nc trek was part of the reason that caused and contributed to the death, however, the 2 xience skypoint stents did not cause or contribute to the death.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key15551526
Report Number2024168-2022-10375
Device Sequence Number16843246
Product Code NIQ
UDI-Device Identifier08717648233111
UDI-Public08717648233111
Combination Product (Y/N)Y
Initial Reporter StateAR
Initial Reporter CountryUS
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/07/2023
Device Model Number1804250-33
Device Catalogue Number1804250-33
Device Lot Number1083141
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/30/2022
Initial Report FDA Received Date10/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/08/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
7 FRENCH GUIDE CATHETER; NC TREK BALLOON; XIENCE SKYPOINT
Outcome Attributed to Adverse Event Required Intervention;
Patient Age83 YR
Patient SexMale
Patient Weight65 KG
-
-