• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDINAL HEALTH GSTRO FEED TBE W/Y PRT 24FR; TUBES, GASTROINTESTINAL (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDINAL HEALTH GSTRO FEED TBE W/Y PRT 24FR; TUBES, GASTROINTESTINAL (AND ACCESSORIES) Back to Search Results
Model Number 8884720247
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
Customer reports: after one months use, there was rupture of the 24fr covidien probe.
 
Additional Manufacturer Narrative
An investigation is currently underway.Upon completion, the results will be forwarded.
 
Additional Manufacturer Narrative
The device history record review of the reported lot number revealed no discrepancy was found related to the reported issue.A sample analysis could not be performed since no photo samples nor actual sample device was received for evaluation.The root cause could not be determined without photos or samples to evaluate.The investigation was carried out with a multifunctional team, and the findings were that all processes and controls were found to be followed according to procedure, including packaging and all inspections performed on the product.No abnormal conditions were found that could trigger the reported condition.The precautions section in item 5 establishes the following: according to the instructions for use precautions section, ¿replace tube at four (4) week intervals to ensure optimum tube patency.¿ a risk assessment was carried out to determine the observed occurrence for this reported condition.Based on the analysis, the observed occurrence of harm (o) is (3, occasional) which is below the range of the expected occurrence of harm (o) (4, probable).The condition reported by the customer was not confirmed, therefore no action plan is deemed required.The current process is running according to product specifications meeting quality acceptance criteria.We will continue monitoring the process for any adverse trends that require immediate attention.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GSTRO FEED TBE W/Y PRT 24FR
Common Device Name
TUBES, GASTROINTESTINAL (AND ACCESSORIES)
Manufacturer (Section D)
CARDINAL HEALTH
777 west street
mansfield MA 02048
Manufacturer (Section G)
CARDINAL HEALTH
calle 9 sur no. 125 cuidad ind
tijuana 22500
MX   22500
Manufacturer Contact
jill saraiva
777 west street
mansfield, MA 02048
5086183640
MDR Report Key15552212
Report Number9612030-2022-03420
Device Sequence Number6158686
Product Code KNT
UDI-Device Identifier10884521007352
UDI-Public10884521007352
Combination Product (Y/N)N
Initial Reporter CountryCO
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number8884720247
Device Catalogue Number8884720247
Device Lot Number2105438164
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/18/2022
Supplement Date Received by Manufacturer09/18/2022
Initial Report FDA Received Date10/06/2022
Supplement Report FDA Received Date10/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-