• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON & CO. (SPARKS) BD BBL¿ MUELLER HINTON II AGAR 150MM PLATE 24PK; CULTURE MEDIA, ANTIMICROBIAL SUSCEPTIBILITY TEST, MUELLER HINTON AGAR/BROTH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO. (SPARKS) BD BBL¿ MUELLER HINTON II AGAR 150MM PLATE 24PK; CULTURE MEDIA, ANTIMICROBIAL SUSCEPTIBILITY TEST, MUELLER HINTON AGAR/BROTH Back to Search Results
Model Number 221800
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/29/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that bd bbl¿ mueller hinton ii agar has seven occurrences of contamination.The product is cloudy and has changed color.The media contamination was noticed upon receiving.No patient impact was reported.The following information was provided by the initial reporter: customer states media is contaminated.Product is cloudy and has changed color.
 
Additional Manufacturer Narrative
A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
 
Additional Manufacturer Narrative
During manufacturing of material 221800, media is formulated and sent through a high temperature short time sterilizer to remove bioburden.The petri dishes are subjected to uv radiation to decrease bioburden.The petri dishes are filled in a positive pressure hepa filtered environment.The filled plates are cooled and immediately wrapped into sleeves to decrease the introduction of microbes.Sleeves are then packaged into cartons and then transferred to a refrigerated truck 2 to 8 degrees celsius (c) for shipment to the distributor.Bd distributors are provided with the storage guidelines for the shipping and handling of bd media of 2 to 8 degrees c in a dark place.The batch history record review for batch 2242624 was satisfactory per internal procedures.The release testing that is performed on this product does include bioburden testing.A sample of plates are incubated at 25 degrees c and at 35 degrees c for approximately 72 hours.All bioburden testing performed on this batch was satisfactory per bd internal procedures.Affected product does not have any sterility claims; the product is tested for bioburden prior to release to ensure that it conforms to product specifications.However, this does not ensure that the end-user will not receive a contaminated plate.The complaint history was reviewed, and no other complaints have been taken on batch 2242624.No retention samples for this batch were available for investigation.Two photos were received.Photo one shows one plate (lot illegible timestamp 1101) with a large colony covering almost half of the plate.Photo two shows part of a sleeve of plates with contamination visible in the plates through the sleeve, no id visible.Bd has identified a contamination trend for this product and the investigation found opportunities for bioburden reduction in the manufacturing process.A capa (corrective and preventative actions) has been initiated and involves implementing additional cleaning events and evaluation of manufacturing procedures focused on in-process bioburden reduction.Additional trainings are planned with an ongoing training review for cleaning processes.Improvement in observation of contamination is expected as the capa progresses.H3 other text : see h.10.
 
Event or Problem Description
It was reported that bd bbl¿ mueller hinton ii agar has seven occurrences of contamination.The product is cloudy and has changed color.The media contamination was noticed upon receiving.No patient impact was reported.The following information was provided by the initial reporter: customer states media is contaminated.Product is cloudy and has changed color.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD BBL¿ MUELLER HINTON II AGAR 150MM PLATE 24PK
Common Device Name
CULTURE MEDIA, ANTIMICROBIAL SUSCEPTIBILITY TEST, MUELLER HINTON AGAR/BROTH
Manufacturer (Section D)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer (Section G)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15552419
Report Number1119779-2022-01279
Device Sequence Number17687174
Product Code JTZ
UDI-Device Identifier10382902218002
UDI-Public10382902218002
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
PREAMENDMENT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date12/14/2022
Device Model Number221800
Device Catalogue Number221800
Device Lot Number2242624
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/03/2022
Supplement Date Received by Manufacturer02/08/2023
Initial Report FDA Received Date10/06/2022
Supplement Report FDA Received Date02/14/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/30/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-