ABBOTT MEDICAL SJM RIGID SADDLE RING; RING, ANNULOPLASTY Back to Search Results
Model Number RIGID SADDLE RING
Medical Device Problem Code Lack of Effect (4065)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/15/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that on (b)(6) 2022, a 30mm rigid saddle ring was chosen for implant.After the device was implanted, a water injection test was performed and showed limited flap movement.The device was then explanted and replaced with another 30mm rigid saddle ring.The patient was reported as stable and normal.The patient remained hemodynamically stable throughout the procedure.There was no clinically significant delay in procedure and no adverse patient effects.
 
Additional Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Additional Manufacturer Narrative
An event of the native valve leaflets having limited movement after the ring was implanted was reported.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.A returned device assessment could not be performed as the was not returned for analysis.The cause of the reported event could not be conclusively determined.There is no indication of a product quality issue with regards to manufacture, design, or labeling.
 
Event or Problem Description
It was reported that on 15 september 2022, a 30mm rigid saddle ring was chosen for implant.After the device was implanted, a water injection test was performed and showed limited flap movement.The device was then explanted and replaced with another 30mm rigid saddle ring.The patient was reported as stable and normal.The patient remained hemodynamically stable throughout the procedure.There was no clinically significant delay in procedure and no adverse patient effects.Subsequent to the previously filed report, additional information was received that the patient's mitral valve leaflets were moving normally during testing.It was also noted that the patient's valve could not close completely.After the valve was explanted, the user tested the leaflets and stated they were moving.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SJM RIGID SADDLE RING
Common Device Name
RING, ANNULOPLASTY
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL PUERTO RICO, INC. REG#2648612
20 b st caguas west park
caguas 00725
*   00725
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key15553206
Report Number2135147-2022-01512
Device Sequence Number6158715
Product Code KRH
UDI-Device Identifier05414734009447
UDI-Public05414734009447
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K042734
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/16/2023
Device Model NumberRIGID SADDLE RING
Device Catalogue NumberRSAR-30
Device Lot Number6461090
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/15/2022
Supplement Date Received by Manufacturer12/15/2022
Initial Report FDA Received Date10/06/2022
Supplement Report FDA Received Date12/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/17/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-