• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. TWINFIX 2.8 TI W/NDL 22-0 ULTRABRAID; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. TWINFIX 2.8 TI W/NDL 22-0 ULTRABRAID; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE Back to Search Results
Model Number 72202067
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Rupture (2208)
Date of Event 09/14/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during a maxillofacial repair surgery, the twinfix anchor broke.All the broken pieces were removed with tweezers.The procedure was successfully completed with a non-significant surgical delay using a back-up device in an additional bone hole.A void was left.No further complications were reported.
 
Additional Manufacturer Narrative
Internal complaint reference (b)(4).
 
Additional Manufacturer Narrative
H10: h3, h6: the reported device was received for evaluation.There was a relationship found between the device and the reported event.A complaint history review found similar reported events.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the anchor found that no burrs, cracks, or contamination is allowed.A certificate of conformance must also be provided.A visual inspection of the returned device found that it is not in its original packaging.The insertion device, fractured anchor and suture material was returned.The suture material is still installed on the insertion device and both have debris on them.The distal tip of the anchor was returned, the proximal end has fractured away and was returned in the insertion device, still attached to the suture material.A review of the customer provided image finds the complaint insertion device with suture material but no anchor visible on the smith+nephew product box.Based on the condition of the product material found during visual inspection, additional material testing is not required.No further risk to the patient is anticipated as a result of the additional bone hole.Since no further harm is anticipated no further clinical assessment is warranted at this time.The complaint was confirmed, and the root cause was associated with unintended use of the device.Factors that could have contributed to the reported event include excessive force on the device, excessive torque on the device, attempted correction of a damaged device, off-axis insertion, improper preparation of the insertion site, or an inadvertent impact event inconsistent with normal use.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TWINFIX 2.8 TI W/NDL 22-0 ULTRABRAID
Common Device Name
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15554068
Report Number1219602-2022-01488
Device Sequence Number11462282
Product Code MBI
UDI-Device Identifier03596010619112
UDI-Public03596010619112
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K9723326
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/20/2026
Device Model Number72202067
Device Catalogue Number72202067
Device Lot Number2068935
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/14/2022
Supplement Date Received by Manufacturer12/20/2022
Initial Report FDA Received Date10/06/2022
Supplement Report FDA Received Date12/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/20/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age35 YR
Patient SexMale
Patient Weight70 KG
-
-