• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US LCS COMPLETE M/B PAT POR STD+; LCS COMPLETE : KNEE PATELLA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US LCS COMPLETE M/B PAT POR STD+; LCS COMPLETE : KNEE PATELLA Back to Search Results
Model Number 1294-09-650
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Code Pain (1994)
Date of Event 09/05/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Lcs complete poly liner and patella revision.Implant currently in situ had wear and delamination on both implant components.Previous operation was complete in 2011.Patient experienced pain due to wear on poly and patella.Revision took place on (b)(6) 2022.New tibial insert and patella put in.Doi: (b)(6) 2011.Dor: (b)(6) 2022.Unknown side.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
Additional information received.The affected side of the patient is the left knee.
 
Event or Problem Description
Additional information received indicated that there was no surgical delay reported during the revision.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected: h6 (the device code was updated from implant bearing wear: polyethylene to implant bearing wear: unknown material.).
 
Additional Manufacturer Narrative
Product complaint (b)(4).Investigation summary: the device associated with this complaint was not returned to depuy synthes for examination.The photo investigation found heavy on the device.Confirming, the reported allegation.Additional monitoring, for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: the device associated with this complaint was not returned to depuy synthes for examination.The photo investigation found heavy wear on the device, confirming the reported allegation.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LCS COMPLETE M/B PAT POR STD+
Common Device Name
LCS COMPLETE : KNEE PATELLA
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI  
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key15554865
Report Number1818910-2022-18392
Device Sequence Number5635100
Product Code NJL
UDI-Device Identifier10603295022657
UDI-Public10603295022657
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 09/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2016
Device Model Number1294-09-650
Device Catalogue Number129409650
Device Lot Number3251832
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/05/2022
Supplement Date Received by Manufacturer10/05/2022
11/17/2022
12/01/2022
12/19/2022
07/19/2023
Initial Report FDA Received Date10/06/2022
Supplement Report FDA Received Date10/10/2022
11/21/2022
12/02/2022
12/22/2022
07/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/04/2011
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DEPUY DUOFIX MBT TRAY SZ 3; LCS COMP RP INSERT STD+ 10MM; LCS COMPLETE FEM POR L STD+; LCS COMPLETE M/B PAT POR STD+
Outcome Attributed to Adverse Event Required Intervention;
Patient Age60 YR
Patient SexFemale
-
-