Model Number 50-746-24-09
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Impaired Healing (2378)
Date of Event 09/22/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a plate broke.It was removed and replaced.
 
Additional Manufacturer Narrative
An investigation was performed using a stereo microscope revealed tensile cracks.Further observation determined there were no indications of material or manufacturing defects.The complaint percentage was calculated, and it is determined that the complaint percentage falls within the design risk limits adhered to at kls se.During the investigation the product lot number identified was reviewed in the device history records.The dhr review showed no discrepancies or anomalies.The investigation results conclude that the root cause for breakages were due to structural failure of the device due to mechanical overload.If further information is obtained that might add value to the contents of the investigation report, an additional follow-up report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL ONE CMF
Common Device Name
PLATE
Manufacturer (Section D)
KARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG
kolbinger strasse 10
muehlheim/donau, 78570
GM  78570
Manufacturer (Section G)
KLS MARTIN L.P.
p.o. box 16369
jacksonville FL 32245
Manufacturer Contact
melissa bachorski
p.o. box 16369
jacksonville, FL 32245
9046417746
MDR Report Key15555705
Report Number9610905-2022-00074
Device Sequence Number6511465
Product Code MQN
UDI-Device Identifier00888118077875
UDI-Public(01)00888118077875
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K032442
Exemption Number2017029
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number50-746-24-09
Device Lot Number33231838
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer10/04/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/22/2022
Supplement Date Received by Manufacturer01/25/2024
Initial Report FDA Received Date10/06/2022
Supplement Report FDA Received Date02/23/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/04/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;