| Model Number |
68HP-750 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
08/22/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a right side device deflation.Device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: ¿ deflation: one opening on posterior side assessed as an unidentified (tear) opening (shell thickness was within specification).Additional observations: ¿ creases are observed on the surface of the shell.¿ white particles observed on the surface of the shell.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a right side device deflation.Device has been explanted and replaced.
|
| |
|