Model Number 7300
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/09/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Hologic is submitting this report in accordance with 21.Cfr part 803.The submission is based upon the currently available information.The submission of this report does not represent a confirmation of device malfunction involving the hologic products in the event described.A device history record (dhr) review was conducted for the reported lot/serial number.The device was released meeting all qa specifications.We are currently unable to establish a relationship between the device and the issue reported.The device involved in this event was not returned for evaluation purposes therefore visual and functional analysis of the product could not be performed.We are unable to confirm a relationship between the device and the issue reported and a definitive root cause for the reported event could not be determined.The information obtained during complaint investigation will be included in our global complaint trending and product surveillance will continue to monitor complaints of this type for adverse trends.If the product is received or additional information is obtained, the investigation will be reopened accordingly per standard operating procedure.There are 2 devices involved in this event: 1222780-2022-00276 and 1222780-2022-00277.
 
Event or Problem Description
It was reported that on (b)(6) 2022 an acessa procedure was performed, when the physician went to use the handpiece, got rf sys error 8048, the physician did a hard reset and then got a cj check error, they powered off/on, and it still did not work.They used a second handpiece and when they went to use it, they noticed that the console temp which should have been at or around 37 and was at 0 celsius so it was unable to be used.The procedure had to be aborted and rescheduled.The patient was under general anesthesia and the physician had already inserted two trocars and the disposable handpiece.No injury to the patient and no additional information available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACESSA PROVU HANDPIECE
Common Device Name
COAGULATOR, LAPAROSCOPIC, UNIPOLAR (AND ACCESORIES)
Manufacturer (Section D)
HOLOGIC, INC
250 campus drive
marlborough MA 01752
Manufacturer (Section G)
HOLOGIC, INC.
250 campus drive
marlborough MA 01752
Manufacturer Contact
ariel lafuente
562 parkway
coyol free zone building b24
san jose 20102
CR   20102
MDR Report Key15556020
Report Number1222780-2022-00277
Device Sequence Number20476884
Product Code HFG
UDI-Device Identifier00854763006140
UDI-Public(01)00854763006140(10)C000091601(17)231026
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K181124
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 10/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/26/2023
Device Model Number7300
Device Lot NumberC000091601
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/13/2022
Initial Report FDA Received Date10/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/15/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ACESSA PROVU HANDPIECE
Outcome Attributed to Adverse Event Other;
Patient SexFemale
-
-