• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GRIFOLS DIAGNOSTIC SOLUTIONS INC. PROCLEIX ULTRIO ELITE ASSAY; HIV-1/HIV-2/HCV/HBV DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GRIFOLS DIAGNOSTIC SOLUTIONS INC. PROCLEIX ULTRIO ELITE ASSAY; HIV-1/HIV-2/HCV/HBV DEVICE Back to Search Results
Lot Number 703564
Medical Device Problem Code False Negative Result (1225)
Health Effect - Clinical Code Hepatitis (1897)
Date of Event 07/14/2022
Type of Reportable Event Serious Injury
Event or Problem Description
On (b)(6) 2022, grifols customer gift of hope (united states of america) reported a case of hbv transfusion transmission infection (tti) for the recipient of a kidney from a beating heart organ donor.The donor of the kidney was tested with serology and the procleix ultrio elite screening assays and they were both nonreactive.There was also a hbv tti event for the heart recipient that is documented under a separate report (2032600-2022-00008).Samples from the donor were tested on (b)(6) 2022 and gave the following results: abbott alinity anti-hbc: nonreactive, abbott alinity hbsag: nonreactive, ultrio elite screening: nonreactive.Ultrio elite ml 703564 was used for testing.The tti was discovered when the patient who received the kidney transplant completed their post-operative follow-up testing (exact date unknown, approximately 4 to 6 weeks post-op) and they were hbv positive.Additional details of the testing were requested but the only information the customer provided was that the transplant team used an hbv pcr test.The customer declined to provide any information from the donor questionnaire, citing confidentiality.The sample tubes used for procleix (k2edta) and serology (preservative and coagulant free) testing were collected at the same time.Grifols reviewed customer reagent and storage and preparation and did not find anything unusual.The ultrio elite run report from the customer site was reviewed by grifols.There were no instrument errors.The run was normal and valid.Calibrators were valid with no unusual values.Internal control results for the sample did not indicate any issues with amplification.The sample was not tested using the ultrio elite discriminatory assays.Review of the device history record (dhr) for ultrio elite ml 703564 was completed.Dhr review confirmed that the master lot performed as expected and met all qc release sensitivity and specificity criteria for both screening and discriminatory assays.A review of grifols complaint records created for ultrio elite hbv sensitivity issues for one year prior to the occurrence date through the present was completed.There is no complaint information that indicates any hbv sensitivity problem with ultrio elite ml 703564 or with the ultrio elite assay in general.An investigation into the root cause of the nonreactive hbv result on ultrio elite is ongoing.Additional information has been requested.The customer had sample available for investigational testing at grifols.Sample was delivered to grifols on (b)(6) 2022 and investigational testing is in progress.An update to this report will be filed with results of the investigational testing once they have been completed.
 
Event or Problem Description
On (b)(6) 2022, (b)(6) (united states of america) reported a case of hbv transfusion transmission infection (tti) for the recipient of a kidney from a beating heart organ donor.The donor of the kidney was tested with serology and the procleix ultrio elite screening assays and they were both nonreactive.There was also a hbv tti event for the heart recipient that is documented under a separate report (2032600-2022-00008).Samples from the donor were tested on (b)(6) 2022 and gave the following results: abbott alinity anti-hbc: nonreactive, abbott alinity hbsag: nonreactive, ultrio elite screening: nonreactive.Ultrio elite ml 703564 was used for testing.The tti was discovered when the patient who received the kidney transplant completed their post-operative follow-up testing (exact date unknown, approximately 4 to 6 weeks post-op) and they were hbv positive.Additional details of the testing were requested but the only information the customer provided was that the transplant team used an hbv pcr test.The customer declined to provide any information from the donor questionnaire, citing confidentiality.The sample tubes used for procleix (b)(4) and serology (preservative and coagulant free) testing were collected at the same time.Grifols reviewed customer reagent and storage and preparation and did not find anything unusual.The ultrio elite run report from the customer site was reviewed by grifols.There were no instrument errors.The run was normal and valid.Calibrators were valid with no unusual values.Internal control results for the sample did not indicate any issues with amplification.The sample was not tested using the ultrio elite discriminatory assays.Review of the device history record (dhr) for ultrio elite ml 703564 was completed.Dhr review confirmed that the master lot performed as expected and met all qc release sensitivity and specificity criteria for both screening and discriminatory assays.A review of grifols complaint records created for ultrio elite hbv sensitivity issues for one year prior to the occurrence date through the present was completed.There is no complaint information that indicates any hbv sensitivity problem with ultrio elite ml 703564 or with the ultrio elite assay in general.An investigation into the root cause of the nonreactive hbv result on ultrio elite is ongoing.Additional information has been requested.The customer had sample available for investigational testing at grifols.Sample was delivered to grifols on (b)(4) 2022 and investigational testing is in progress.An update to this report will be filed with results of the investigational testing once they have been completed.Final report: the customer provided the donor plasma sample to grifols for investigational testing on (b)(6) 2022.Hbv quantitative testing was completed on (b)(6) 2022.Investigational testing was performed by testing the sample using the hologic aptima hbv quant assay (aptima qhbv).The testing result was "not detected".Per the aptima qhbv package insert (aw-15644, rev.002), the limit of detection (lod) of the assay using the 3rd who international standard for hbv dba (nibsc 10/264, genotype a) is 4.8 iu/ml for plasma and 5.9 iu/ml for serum.Based on the aptima qhbv "not detected" result and the lod indicated in the aptima qhbv package insert, if hbv is present in the plasma sample it is at a concentration below 4.8 iu/ml.The ultrio elite package insert (gdss-ifu-000006 v7.0) shows 95% detection probability of 4.3 iu/ml with 95% fiducial limits of 3.8 - 5.0 iu/ml for the hbv who (97/750) international standard on the ultrio elite screening assay.Based on the results of hbv quantitative testing, the hbv concentration of the sample is at or below the ultrio elite screening hbv lod and therefore would not be expected to give consistently reactive results.The root cause of the nonreactive ultrio elite result is low titer sample and the ultrio elite assay is performing as designed.No further information is expected, this is the final report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROCLEIX ULTRIO ELITE ASSAY
Common Device Name
HIV-1/HIV-2/HCV/HBV DEVICE
Manufacturer (Section D)
GRIFOLS DIAGNOSTIC SOLUTIONS INC.
4560 horton street
emeryville CA 94608
Manufacturer Contact
amanda doe
10808 willow court
san diego, CA 92127
8582020852
MDR Report Key15558448
Report Number2032600-2022-00007
Device Sequence Number12533477
Product Code QHO
UDI-Device Identifier00859882007603
UDI-Public00859882007603
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
BL125652
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Lot Number703564
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/09/2022
Supplement Date Received by Manufacturer09/09/2022
Initial Report FDA Received Date10/07/2022
Supplement Report FDA Received Date11/04/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-