GLAUKOS CORPORATION ISTENT INFINITE TRABECULAR MICRO-BYPASS SYSTEM; AQUEOUS SHUNT Back to Search Results
Model Number IS3-WW
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Cataract (1766); Iritis (1940)
Date of Event 09/21/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device was not available; therefore, product testing on the actual device could not be performed.The device history record review of the manufacturing lot was performed and there were no non-conformities found to be related to the reported event.A review of the device labeling was completed.Iritis identified in the labeling as known inherent risk of trabecular micro-bypass aqueous shunt procedure.The ifu adequately provides instructions, precautions, and warnings for the proper use and handling of the device.An adverse event appears to have occurred, but does not appear to have been a problem with the device or the way it was used.The reported event is an established risk associated with use of the device, which is clearly specified in the product's labeling.Based on the information received, the root cause of the reported event could not be conclusively identified.Mfr# reference: (b)(4).
 
Event or Problem Description
It was reported that following an istent infinite trabecular microbypass system procedure, the patient presented one (1) week postoperatively with mild iritis associated with 1+ cell noted in the anterior chamber, which was treated with medication.The report noted the event as ongoing.Additional information has been requested.
 
Event or Problem Description
The following additional information was received.Per report, the patient presented with mild transillumination of mid periphery and mild cortical changes of the lens occurring on (b)(6) 2022.The investigator indicated both ongoing events as non-serious.There was no report of any interventions required.
 
Additional Manufacturer Narrative
Mfr# reference: (b)(4).
 
Additional Manufacturer Narrative
Additional information: b5, b6, g3, g4.Mfr# reference: (b)(4).
 
Event or Problem Description
Through follow-up, the following information was provided.The patient''s most recent examination noted no further interventions required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ISTENT INFINITE TRABECULAR MICRO-BYPASS SYSTEM
Common Device Name
AQUEOUS SHUNT
Manufacturer (Section D)
GLAUKOS CORPORATION
229 avenida fabricante
san clemente CA 92672
Manufacturer (Section G)
GLAUKOS CORPORATION
229 avenida fabricante
san clemente CA 92672
Manufacturer Contact
brittany massey
229 avenida fabricante
san clemente, CA 92672
949367960
MDR Report Key15559729
Report Number2032546-2022-00110
Device Sequence Number7073711
Product Code KYF
UDI-Device Identifier00853704002470
UDI-Public0100853704002470
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K220032
Number of Events Summarized123
Summary Report (Y/N)Y
Device Implanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Study,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/31/2023
Device Model NumberIS3-WW
Device Lot Number112718
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Fifteen-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Not provided
Initial Report FDA Received Date10/07/2022
Supplement Report FDA Received Date11/04/2022
01/16/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/30/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age55 YR