DEPUY IRELAND - 9616671 ATTUNE REV LPS INSRT XSM 12MM; KNEE TIBIAL INSERT Back to Search Results
Model Number 1517-60-212
Medical Device Problem Code Difficult to Open or Remove Packaging Material (2922)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/22/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the lps insert and the sterile paper packaging wouldn't remove the sterile layer to allow the implant to be opened correctly.Tried a different corner to get the paper to release, but all four corners were stuck fast and ended up tearing between the layers of packaging paper.
 
Event or Problem Description
Additional information received.The affected side of the patient was the left knee.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).Investigation summary: the device associated with this report was not returned.Based on the visual analysis of the provided photographic evidence, it was possible to confirm that tyvek lid of was not possible to be properly retrieved from the adhesive of the sterile package.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: nr-0190177 was launched in depuy synthes quality systems,  upon review of the audit trail, it was found that batch jp9016 was sent for sterilization twice due to damaged and dirty condition of unit cartons.They were dispositioned to have the unit cartons and unit carton labels replaced.The non-standard rework activity did not call for sterilization of the product upon completion, as the sterile barrier was not breached as part of the rework activity.However, upon completion of the rework activity and systematic release of the product, it was inadvertently sent through the sterilization process a second time.
 
Additional Manufacturer Narrative
H7 and h9: recall information provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE REV LPS INSRT XSM 12MM
Common Device Name
KNEE TIBIAL INSERT
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY RAYNHAM, A DIV. OF DEPUY ORTHO 1219655
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key15561779
Report Number1818910-2022-19782
Device Sequence Number5684857
Product Code KRO
UDI-Device Identifier10603295490968
UDI-Public10603295490968
Combination Product (Y/N)N
Initial Reporter CountryNZ
PMA/510(K) Number
K191779
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 10/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1517-60-212
Device Catalogue Number151760212
Device Lot NumberJP9016
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/22/2022
Supplement Date Received by Manufacturer10/20/2022
01/26/2023
02/28/2024
Initial Report FDA Received Date10/07/2022
Supplement Report FDA Received Date10/24/2022
01/26/2023
03/21/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/16/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
1818910-02/09/2023-001-R
Patient Sequence Number1