• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER EDGEPLUS VALVED ENTRY SYSTEM; CANNULA, TROCAR, OPHTHALMIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER EDGEPLUS VALVED ENTRY SYSTEM; CANNULA, TROCAR, OPHTHALMIC Back to Search Results
Catalog Number 8065751658
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/14/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A nurse reported that during a vitrectomy surgery an ophthalmic trocar was leaking when it removed from the stab knives.Surgery was completed by using an alternate product and no patient harm was reported.
 
Additional Manufacturer Narrative
The returned samples were visually inspected and were found to be non-conforming with torn septums.Leak testing could not be performed due to the damage of the samples.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The photo attached to the parent complaint was reviewed by the manufacturing site.The photo is of a blister label, the reported product and lot information is confirmed.The video attached to the parent complaint was reviewed by the manufacturing site.The video consists of an eye surgery, the reported complaint was not confirmed.The exact root cause for this complaint is unknown; the damaged condition of the valves could have contributed to the reported event; how and when the valves became damaged cannot be determined from the evaluation performed.The exact root cause for this complaint is unknown therefore specific action for this complaint cannot be taken.An acceptance quality inspection is performed to ensure product meets release acceptance criteria.This inspection includes evaluation for damaged valves.Complaints are reviewed and monitored at regular intervals to evaluate effectiveness of actions taken.No adverse trends have been observed associated with the reported product and event.No further actions are required.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDGEPLUS VALVED ENTRY SYSTEM
Common Device Name
CANNULA, TROCAR, OPHTHALMIC
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15561855
Report Number2028159-2022-01420
Device Sequence Number17785868
Product Code NGY
UDI-Device Identifier00380657516582
UDI-Public00380657516582
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 01/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2023
Device Catalogue Number8065751658
Device Lot Number13VJ0D
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/15/2022
Supplement Date Received by Manufacturer12/18/2022
Initial Report FDA Received Date10/07/2022
Supplement Report FDA Received Date01/16/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/22/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-