Product complaint #: (b)(4).A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Investigational summary: the product was returned to ethicon for evaluation.Visual inspection evaluation was conducted on the returned device.Visual analysis of the returned sample determined that one opened overwrap with an empty labeled winding former one detached needle and one suture piece that pertains to product code 1162h was returned for analysis.Upon visual analysis of the returned sample, the swage and attachment area were noted to be as expected.No suture remnant could be observed in the barrel hole.During the visual inspection of the suture, marks on the surface were noted due to the handling during use.The insertion end was not observed due to the needle was observed to be cut from the strand possibly caused by a surgical instrument.As per returned conditions of the sample no functional test was performed.The condition of the sample received indicates improper handling of the device.The condition of the sample received indicates improper handling of the device.It should be noted that as part of our quality process, each batch is randomly inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Related reports: 2210968-2022-06483.
|