| Model Number |
115-150 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Codes |
Capsular Contracture (1761); Seroma (2069)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Patient reported no complaint against the right device.Physician later reported right side rupture.The events of "isoechoic fluid collection between implant bag and capsule", "focal detachment of implant c several folding of implant bag", "sever capsule thickening c irregular implant contour" and "double capsule c contracture" baker grade unknown were reported via us report.The event of "several cysts and hypoechoic nodules" is not device related.Device was explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.The events of capsular contracture and seroma are physiological complications and analysis of the device generally does not assist allergan in determining a probable cause for these events.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture baker grade unknown, rupture, seroma late, crease folding, double capsule, malposition lump nodule, cyst.
|
| |
|
Event or Problem Description
|
|
Patient reported no complaint against the right device.Physician later reported right side rupture.The events of "isoechoic fluid collection between implant bag and capsule", "focal detachment of implant c several folding of implant bag", "sever capsule thickening c irregular implant contour" and "double capsule c contracture" baker grade unknown were reported via us report.The event of "several cysts and hypoechoic nodules" is not device related.Device was explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: rupture: one opening observed on posterior side assessed as fold flaw opening.Seroma-late: unable to observe as it is a medical event not related to the device.Capsular contracture: unable to observe as it is a medical event not related to the device.Double capsule: unable to observe as it is a medical event not related to the device.Crease/folding of implant: creases were observed.Malposition: unable to observe as it is a medical event not related to the device.Lump/nodule-ndr: not applicable as the event is not related to the device cyst-ndr: not applicable as the event is not related to the device.Additional observations: creases were observed.Wear abrasion observed.Blank particles observed on the surface of the shell.Stress marks observed.
|
| |
|