ALLERGAN (COSTA RICA) STYLE 115 SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Model Number 115-150
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Codes Capsular Contracture (1761); Seroma (2069)
Type of Reportable Event Serious Injury
Event or Problem Description
Patient reported no complaint against the right device.Physician later reported right side rupture.The events of "isoechoic fluid collection between implant bag and capsule", "focal detachment of implant c several folding of implant bag", "sever capsule thickening c irregular implant contour" and "double capsule c contracture" baker grade unknown were reported via us report.The event of "several cysts and hypoechoic nodules" is not device related.Device was explanted.
 
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.The events of capsular contracture and seroma are physiological complications and analysis of the device generally does not assist allergan in determining a probable cause for these events.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture baker grade unknown, rupture, seroma late, crease folding, double capsule, malposition lump nodule, cyst.
 
Event or Problem Description
Patient reported no complaint against the right device.Physician later reported right side rupture.The events of "isoechoic fluid collection between implant bag and capsule", "focal detachment of implant c several folding of implant bag", "sever capsule thickening c irregular implant contour" and "double capsule c contracture" baker grade unknown were reported via us report.The event of "several cysts and hypoechoic nodules" is not device related.Device was explanted.
 
Additional Manufacturer Narrative
Device evaluation: based on the device analysis grid, the assessments of the complaint are: rupture: one opening observed on posterior side assessed as fold flaw opening.Seroma-late: unable to observe as it is a medical event not related to the device.Capsular contracture: unable to observe as it is a medical event not related to the device.Double capsule: unable to observe as it is a medical event not related to the device.Crease/folding of implant: creases were observed.Malposition: unable to observe as it is a medical event not related to the device.Lump/nodule-ndr: not applicable as the event is not related to the device cyst-ndr: not applicable as the event is not related to the device.Additional observations: creases were observed.Wear abrasion observed.Blank particles observed on the surface of the shell.Stress marks observed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 115 SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer Contact
terry ingram
12331-a riata trace parkway
building 3
austin, TX 78727
8479366324
MDR Report Key15562067
Report Number9617229-2022-18212
Device Sequence Number20255094
Product Code FTR
UDI-Device Identifier10888628002302
UDI-Public10888628002302
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Foreign,Consumer,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/20/2020
Device Model Number115-150
Device Catalogue Number115-150
Device Lot Number2744219
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/13/2022
Supplement Date Received by Manufacturer01/02/2023
Initial Report FDA Received Date10/07/2022
Supplement Report FDA Received Date01/17/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/20/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
Patient Weight55 KG
Patient RaceAsian
-
-