Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
Product complaint # (b)(4).Investigation summary: the device associated with this complaint was not returned.Photo evidence and x-rays provided were reviewed and found no implant issue.The reported condition was not confirmed.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: a search of the depuy nonconformance (nc) quality system found no nc¿s associated with this product code 960102 /lot code 2335515 combination.Based on the inability to find any nc¿s against the provided product code 960102 /lot code 2335515 combination, it is reasonable to conclude that there are no anomalies with regard to manufacturing or inspection contained in the device history records that would contribute to the reported event.
|