| Catalog Number |
338.25P |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/30/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A nurse reported that during vitrectomy an ophthalmic scraper was opened inclined to the left side, and was not in the desired position in the fully open state.Procedure was completed successfully.There was no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record (dhr) traceable to the reported lot number, indicates that the product was processed and released according to the product¿s acceptance criteria.Sample was not received, by the investigation site for evaluation.A sample was not received, at the manufacturing site.Therefore, the root cause for the customer complaint issue cannot be determined.The exact root cause for the customers reported event is unknown.Therefore, specific action cannot be taken.Complaints are reviewed and monitored at regular intervals for adverse trends.No adverse trends have been observed, associated with the reported product and event.No action has been identified for this reported event.The manufacturer internal reference number is: (b)(4).
|
| |
|